Comment 'Answer: It looks lik...' to 'Catheter Stuck in Radial artery!'
Comment to Catheter Stuck in Radial artery!
-
-
·
F Idris
✓
- ·
Answer:
It looks like the guide catheter was stuck in the radial artery. Spasm could be a factor and sometime the edge of the guide catheter (which is a bit sharp) can make advancement hard.
Tip:
So, all the above suggestions are reasonable and worth trying. However, another technique you can do is to advance an interventional guide-wire into the guide catheter, then advance a balloon over it and make part of it inside the catheter and the other outside it, then inflate that balloon to a low pressure, after that you may be able to advance the catheter and the balloon together (Watch attached video).
-
-
·
Khaled Sleik
Scrubdin Team✓
-
·
In reply to F Idris
F
- ·
Inflating the balloon and advancing the guide is a risk to scrape and damage the vascular system .
AUCATH guide Hydrophilic works 100% of the time
-
·
Khaled Sleik
-
-
·
F Idris
✓
-
·
In reply to Khaled Sleik
- ·
-
·
F Idris
-
·
F Idris